Array-optimized composite pulse for excellent whole-brain homogeneity in high-field MRI.
Journal
  Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine.
Citation
  Magn Reson Med. 57(3):470-4
Publication date
  2007 Mar
Authors
  Collins CM
Wang Z
Mao W
Fang J
Liu W
Smith MB
Investigators
  Christopher Michael Collins
Jidong Fang
Wenlei Liu
Michael B. Smith
Grant agencies
  National Center for Research Resources
National Institute for Biomedical Imaging and Bioengineering
Grants
  NCRR P41RR008079
NCRR P41RR16105
NIBIB R01EB000454
NIBIB R01EB000895
Abstract
  A number of methods to improve excitation homogeneity in high-field MRI have been proposed, and some of these methods rely on separate control of radiofrequency (RF) coils in a transmit array. In this work we combine accurate RF field calculations and the Bloch equation to demonstrate that by using a sequence of pulses with individually optimized current distributions (i.e., an array-optimized composite pulse), one can achieve remarkably homogeneous distributions of available signal intensity over the entire brain volume. This homogeneity is greater than that achievable using the same transmit array to produce either a single optimized (or RF shimmed) pulse or a single RF shimmed field distribution in a standard 90x-90y composite pulse arrangement. Simulations indicate that with a very simple array-optimized composite pulse, excellent whole-brain excitation homogeneity can be achieved at up to 600 MHz.